Hot Laboratory - SPECT

Print
Edit it in Health Facility Briefing System (HFBS)
Edit this room data or print a PDF using the HFBS - Health Facility Briefing System. Customise this room data for your project, or add it to an existing health facility brief in HFBS Briefing Module. Edit Briefing
Room Layout Sheet
Download a PDF Room Layout Sheet (RLS) for this standard component. Free View
Other Files
Buy an editable AutoCAD DWG Room Layout Sheet (RLS) for this standard component. $2.99
Online Room Data Sheet
Room Code HTLB-I
Net Room Area Briefed 12
Hours of Operation 8am to 6pm, daily
Occupancy 1
Description / Special Requirements The Hot Laboratory is an area for storage and preparation of radiopharmaceuticals for use in imaging procedures. The area will include a preparation bench, cupboards and sink with touch free taps. Space will be required for dose calibration, record keeping and quality assurance activities. If pre-prepared materials are used, storage and calculation areas may be considerably smaller than for on-site preparation. Radionuclide storage areas and waste storage within the space may require sheilding, including benches and cupboards. The Hot Laboratory will be located with ready access to Nuclear Imaging Rooms and dosing areas. Additional Design Considerations - Radiation shielding to be advised by Radiation Consultant - The floors and walls should be constructed of a material that is easily decontaminated, with no gaps or crevices - Vents and traps for radioactive gases should be provided if such are used
Amendment 24.10.10, - First Draft 02.05.11, - FE remarks updated 17.02.2026 Room name updated and RDS updated. Room area updated
ID Fabric Material Finish Specification Selection Remarks
1000 Floor finish Vinyl Standard slip resistant Seamless, coved
3466 Skirting Vinyl Prefinished Floor vinyl coved, 125mmH
3000 Wall finish Paint Acrylic, washable
5010 Ceiling Plasterboard Paint, washable Flush set, suspended
3022 Wall finish Resinate, 6mm Prefinished Splashback above bench
6021 Cornice Tape set Paint Set between plasterboard wall and ceiling
8010 Door protection Composite Prefinished PVC Protection plate to 900mm AFFL
8040 Door protection Composite Prefinished PVC To door frame
1709 Doors Solid Core Laminate Single leaf, with lead shielding 920mm clear opening, lockable
ID Description Category Group Qty Selection / Remarks
4460 Cupboard: wall mounted Furniture/ Fitting 1 1 over bench, adjustable shelves
6152 Dispenser: disposable glove Furniture/ Fitting 2 1
6154 Dispenser: paper/ hand towel Furniture/ Fitting 2 1 to sink
6156 Dispenser: soap Furniture/ Fitting 2 1 to sink
8665 Syringe carrier: lead lined MEQ - Nuclear Medicine 3 2
8675 Shield: lead, brick MEQ - Nuclear Medicine 3 8
8664 Survey meter: scintillation MEQ - Nuclear Medicine 3 1
8611 Waste bin: radioactive, 20 litres MEQ - Nuclear Medicine 3 1
8676 Shield: syringe, 3cc MEQ - Nuclear Medicine 3 2 for TC99
8677 Shield: syringe, 5cc MEQ - Nuclear Medicine 3 2 for TC99
8831 Shield: vial, lead MEQ - Nuclear Medicine 3 5
8986 Shield: syringe, 2cc MEQ - Nuclear Medicine 3 2 for TC99
9061 Tongs: radiation protection MEQ - Nuclear Medicine 3 2
9062 Survey meter: contamination, G-M detector MEQ - Nuclear Medicine 3 1
8955 Waste bin: radioactive, beta, 20 litres MEQ - Nuclear Medicine 3 1
8663 Sharps disposal: lead lined MEQ - Nuclear Medicine 3 1 in radiation shielded cupboard
9064 Dosimeter: pocket MEQ - Nuclear Medicine 3 1
1010 Bench: stainless steel Metalwork 1 1
8671 Cabinet: storage, lead lined, w shelves & drawers MEQ - Nuclear Medicine 2 1
8680 Shield: lead, benchtop MEQ - Nuclear Medicine 3 1
9065 Kit: emergency, radioactivity spillage, on trolley MEQ - Nuclear Medicine 3 1
9188 Shield: syringe, 5cc, acrylic, beta pharmaceuticals MEQ - Nuclear Medicine 3 1
9816 Forceps: vial, curved MEQ - Nuclear Medicine 3 5
9819 Survey meter: ion chamber MEQ - Nuclear Medicine 3 2
9857 Shield: vial, Y-90 MEQ - Miscellaneous 3 1
10157 Allowance: hot laboratory, SPECT MEQ - Nuclear Medicine 3 1 to include additional syringe shields, vial shields, syringe carriers, tongs... etc, details and quantity depend on operational requirements
10158 Allowance: QC & calibration sources MEQ - Nuclear Medicine 3 1 as required, depending on operational requirements
ID Description Grp Qty Ele Data CdW HtW WmW Tap Dns Gas Selection / Remarks
9140 Well counter: wipe testing, lead lined 3 1 yes opt ___ ___ ___ ___ ___ ___ mounted on hot cell
9142 Dose calibrator: multipurpose 3 1 yes opt ___ ___ ___ ___ ___ ___
9166 Radiation area monitor: digital, wall mounted 2 2 yes opt ___ ___ ___ ___ ___ ___ one inside the room and one outside the room
9132 Cabinet: storage, lead lined, sink and waste 2 1 ___ ___ yes ___ yes yes yes ___
9867 Laminar flow hood: radioisotope, with base cabinet, 2000mm 2 1 yes ___ ___ ___ ___ ___ ___ ___
9056 Refrigerator: drugs, under-bench, 150L 3 1 yes yes ___ ___ ___ ___ ___ ___
7300 Camera: CCTV security 1 1 yes ___ ___ ___ ___ ___ ___ ___
ID Description Service Category Qty Selection / Remarks
5000 Airconditioning HVAC
2189 General LED Lighting
6055 Special: task light, built in Lighting under O/H cupboards
2577 PO: Double Power 2 to walls
2578 PO: Emergency power, double Power 3 1 - refrigerator; 1 - dose calibrator/ well counter bench side; 1 - dose calibrator/ well counter behind the hood
12011 Radiation shielding Shielding to walls, floor, ceiling, doors & glazing as specified by Radiation Consultant
4010 Cold water: outlet Hydraulics 1 to lead-lined sink and waste cabinet (refer to manufacturer specifications)
4030 Drainage Hydraulics 1 to lead-lined sink and waste cabinet (refer to manufacturer specifications)
4000 Hot water: outlet Hydraulics 1 to lead-lined sink and waste cabinet (refer to manufacturer specifications)
1005 Voice / Data outlet: single Communications 1 1 - refrigerator
2579 PO: Emergency power, single Power 5 2- wall mounted radiation area monitor; 3 - radioisotope hood (from top)
7022 Compressed air Medical Gases 1 to radioisotope hood (optional or gas)
5019 Exhaust HVAC 1 to radioisotope hood
10020 Card access Security 1
2233 Direct connection: Emergency power Power 1 to CCTV camera
1006 Voice / Data outlet: double Communications 2 1 - dose calibrator/ well counter bench side; 1 - dose calibrator/ well counter inside the hood
0 0