| Room Code | MRI-OP-42-I |
| Net Room Area Briefed | 42 |
| Hours of Operation | 8am to 10pm, daily |
| Occupancy | 1 patient; up to 4 staff assisting |
| Description / Special Requirements | Room for MRI scanning procedures, with adjacent control room and computer room. A staff handwashin basin will be located outside the room in close proximity. Fixtures and fittings to be non magnetic. Body Protected electrical area Acoutsics: acoustic privacy required for ambient noise emitted by scanning equipment Special Design Requirements: - The MRI suite shall be planned to conform to staged access and control zones complying with local authority requirements to prevent unscreened persons accessing the 5-gauass volume around the MRI scanner and to minimise radiofrequency interference from other equipment - Electromagnetic sheilding (radio frequency enclosure) as advised by RF Shielding Consultant - Structural assessment may be required to ensure the floor is designed to acommodate the weight of the equipment - Floor to ceiling cable duct with removable covers may be required - Underfloor cable trays between Control area and MRI table and floor to ceiling duct may be required - Warning lights outside each entry door - Power to be on emergency supply - all power to MRI to pass through MRI isolation transformer - Installation of the MRI including services requirements such as cryogen venting, emergency exhaust, passive pressure relief systems, shall conform to local authority requirements and manufacturer's specifications |
| Amendment | 25.09.2025 First Draft |
| ID | Fabric | Material | Finish | Specification | Selection | Remarks |
| 2000 | Skirting | Vinyl | Prefinished | Floor vinyl coved, 150mmH | ||
| 3000 | Wall finish | Paint | Acrylic, washable | |||
| 5010 | Ceiling | Plasterboard | Paint, washable | Flush set, suspended | at 3000mm height | |
| 1000 | Floor finish | Vinyl | Standard slip resistant | Seamless, coved | ||
| 2767 | Cornice | Plasterboard | Paint | Square set | ||
| 9110 | Doors | Solid Core | Paint | 1 1/2 leaf | patient entry/ exit; 1400mm clear opening | |
| 6345 | Observation | RF glass | Clear | Internal window | to control room |
| ID | Description | Category | Group | Qty | Selection / Remarks |
| 8417 | Air flowmeter: MRI | MEQ - General | 3 | 1 | to patient services panel; non-magnetic |
| 9481 | Board: patient transfer, MRI | MEQ - General | 3 | 1 | to be placed in the store |
| 8642 | Bracket: suction bottle, MRI | MEQ - General | 2 | 2 | |
| 8651 | Cabinet: storage, MRI coils | MEQ - Imaging | 3 | part of MRI system configurations | |
| 8419 | Footstool: MRI | MEQ - Furniture | 3 | 1 | |
| 8423 | IV pole: mobile, MRI | MEQ - General | 3 | 1 | |
| 8424 | Suction adapter: MRI | MEQ - General | 3 | 2 | to patient services panel; non-magnetic |
| 9183 | Trolley: multi-purpose, MRI | MEQ - Furniture | 3 | 1 | |
| 8418 | Oxygen flowmeter: MRI | MEQ - General | 3 | 2 | to patient services panel; non-magnetic |
| 2926 | Dispenser: hand sanitiser | Furniture/ Fitting | 2 | 1 | |
| 8321 | Suction bottle | MEQ - General | 3 | 2 | |
| 8421 | Wheelchair: MRI | MEQ - General | 3 | 1 | or can be kept in store |
| ID | Description | Grp | Qty | Ele | Data | CdW | HtW | WmW | Tap | Dns | Gas | Selection / Remarks |
| 7300 | Camera: CCTV security | 1 | yes | ___ | ___ | ___ | ___ | ___ | ___ | ___ | part of MRI system configurations; to view patient, optional | |
| 8777 | Monitor: physiologic, MRI, on stand | 3 | 1 | yes | opt | ___ | ___ | ___ | ___ | ___ | ___ | or can be kept in store |
| 9007 | Ventilator: adult/ paediatric, MRI, on stand | 3 | 1 | yes | opt | ___ | ___ | ___ | ___ | ___ | ___ | or can be kept in store |
| 9052 | Anaesthesia unit: MRI | 3 | 1 | yes | opt | ___ | ___ | ___ | ___ | ___ | ___ | or can be kept in store |
| 9111 | Injector: contrast media, mobile, dual head, MRI | 3 | 1 | yes | yes | ___ | ___ | ___ | ___ | ___ | ___ | |
| 10054 | Scanning unit: MRI, open, 1.2-tesla | 2 | 1 | yes | yes | yes | ___ | ___ | ___ | yes | ___ | number of testla can be as per operational requirements; with patient table and control console |
| ID | Description | Service Category | Qty | Selection / Remarks |
| 1006 | Voice / Data outlet: double | Communications | 2 | 1 - to patient services panel; 1 - wall |
| 1040 | Intercom | Communications | 1 | speakers and mic are mounted to the gantry |
| 5000 | Airconditioning | HVAC | ||
| 6001 | General: colour corrected | Lighting | ||
| 6110 | Downlights: dimmable | Lighting | ||
| 7000 | Oxygen (O2) | Medical Gases | 2 | patient services panel; non magnetic outlet |
| 7010 | Nitrous Oxide (N2O) | Medical Gases | 1 | patient services panel; non magnetic outlet |
| 7040 | Suction | Medical Gases | 2 | patient services panel; non magnetic outlet |
| 7070 | Scavenge | Medical Gases | 1 | patient services panel; non magnetic outlet |
| 8080 | Emergency call | Nurse Call | 1 | + indicator button & light |
| 2579 | PO: Emergency power, single | Power | 4 | patient services panel |
| 2578 | PO: Emergency power, double | Power | 4 | walls |
| 9040 | Body protected | Power | ||
| 12020 | Faraday cage | Shielding | to walls, floor, ceiling, doors & glazing as specified by Radiation Consultant | |
| 8010 | Staff/ Nurse assist call | Nurse Call | 1 | + indicator button & light |
| 7020 | Medical Air (MA) | Medical Gases | 1 | patient services panel; non magnetic outlet |
| 2502 | Clean Agent Gas Suppression System | Fire Services | NOVEC 1230 | |
| 2411 | Special: "MRI In Use" light | Lighting | 1 | to outside of room entry |
| 6813 | Wall panel, Removable, Laminate finish | Services Panels | 1 | |
| 2473 | Power over Ethernet (PoE) | Power | 2 | to CCTV camera (as applicable) |
| 2576 | PO: Cleaner | Power | 1 |