Room Code | PETCR-I |
Net Room Area Briefed | 12 |
Hours of Operation | Available 24 Hours |
Occupancy | Up to 3 staff |
Description / Special Requirements | Room adjacent to CT Scanning Room/s with direct visibility to the patient's full body for remote control of equipment, and review and reporting of procedure images. The Control room may serve two procedure rooms. Acoustics: Acoustic privacy is required Additional Design Considerations: - Radiation shielding as advised by Radiation Consultant - Cable ducts from CT Scanning and Computer room may be required |
Amendment | 25.06.2019 First Draft 03.09.2023 RDS updated |
ID | Fabric | Material | Finish | Specification | Selection | Remarks |
2000 | Skirting | Vinyl | Prefinished | Floor vinyl coved, 150mmH | ||
3000 | Wall finish | Paint | Acrylic, washable | |||
1219 | Door Protection | Composite | Prefinished PVC | Kickplate, 300mm high | ||
9001 | Doors | Solid Core | Paint | Single leaf | staff entry/ exit; 920mm clear opening | |
4159 | Cornice | Aluminium | Powdercoat | Angle flush against wall, in line with ceiling grid | ||
5062 | Ceiling | Acoustic | Prefinished | Drop-in tiles, 600x600 mm | or set plasterboard ceiling if scanning room is provided within an operating environment | |
1000 | Floor finish | Vinyl | Standard slip resistant | Seamless, coved | ||
7025 | Observation | Lead glass | Clear | Internal window | to procedure room |
ID | Description | Category | Group | Qty | Selection / Remarks |
2100 | Chair: ergonomic, office | Furniture/ Fitting | 3 | 2 | |
5598 | Shelving unit: adjustable, full height | Furniture/ Fittings | 1 | 1 | manuals, educational material |
9100 | Hook: heavy duty | Metalwork | 1 | 2 | for lead aprons |
1050 | Bench: laminate | Furniture/ Fitting | 1 | 1 |
ID | Description | Grp | Qty | Ele | Data | CdW | HtW | WmW | Tap | Dns | Gas | Selection / Remarks |
11400 | Computer: PC & monitor | 3 | 1 | yes | yes | ___ | ___ | ___ | ___ | ___ | ___ | |
54510 | Telephone: handset, standard | 3 | 1 | ___ | yes | ___ | ___ | ___ | ___ | ___ | ___ | |
9092 | Control console | 2 | 1 | yes | yes | ___ | ___ | ___ | ___ | ___ | ___ | part of PET/ CT scanning unit |
35860 | Monitor: CCTV | 3 | 1 | yes | yes | ___ | ___ | ___ | ___ | ___ | ___ | part of PET/ CT scanning unit |
ID | Description | Service Category | Qty | Selection / Remarks |
1006 | Voice / Data outlet: double | Communications | 5 | to control console and computers |
1040 | Intercom | Communications | 1 | with handsfree microphone to PET/ CT scanning room |
5000 | Airconditioning | HVAC | ||
2189 | General LED | Lighting | ||
12011 | Radiation shielding | Shielding | to wall, door & glazing shared with PET/ CT scanning room as specified by Radiation Consultant | |
6076 | Special: 'X-ray In Use' lights | Lighting | 1 | over door to procedure room |
2578 | PO: Emergency power, double | Power | 1 | to wall |
2726 | PO: UPS, double | Power | 10 | to control console and computers |