Room Code WS-BM-20-I
Net Room Area Briefed 20
Hours of Operation Available 24 Hours
Occupancy 1 - 2
Description / Special Requirements The Biomedical Workshop provides an area for the repair, maintenance and calibration of biomedical equipment, used in the hospital.
Amendment 23.12.2024 First Draft
ID Fabric Material Finish Specification Selection Remarks
1000 Floor finish Vinyl Standard slip resistant Seamless, coved
3000 Wall finish Paint Acrylic, washable
4020 Wall protection Composite Prefinished PVC Crash rail at 900mm AFFL
9127 Doors Solid Core/ Safety glass Paint 1 1/2 leaf, observation panel 1400mm clear opening, larger leaf to be 920mm clear opening
8010 Door protection Composite Prefinished PVC Protection plate to 900mm AFFL
3060 Wall finish Vinyl Prefinished Splashback to basin & above bench
4159 Cornice Aluminium Powdercoat Angle flush against wall, in line with ceiling grid
5062 Ceiling Acoustic Prefinished Drop-in tiles, 600x600 mm
8040 Door protection Composite Prefinished PVC To door frame
2000 Skirting Vinyl Prefinished Floor vinyl coved, 150mmH
5007 Ceiling Plasterboard Paint Bulkhead
ID Description Category Group Qty Selection / Remarks
1050 Bench: laminate Furniture/ Fitting 1 1
1950 Chair: ergonomic, high Furniture/ Fitting 3 2
4420 Cupboard: under bench, lockable Furniture/ Fitting 1 1
4470 Cupboard: wall mounted, lockable Furniture/ Fitting 1 1 over bench
6152 Dispenser: disposable glove Furniture/ Fitting 2 1
6154 Dispenser: paper/ hand towel Furniture/ Fitting 2 1 to basin
6158 Dispenser: soap antiseptic Furniture/ Fitting 2 1 to basin
8879 Multi-meter: digital MEQ - General 3 1 qty depends on operational requirements
9116 Lux meter MEQ - General 3 1 qty depends on operational requirements
8880 Toolkit: repair, biomedical MEQ - General 3 2 qty depends on operational requirements
10061 Waste bin: General, 20 Litres Furniture/ Fitting 3 1
ID Description Grp Qty Ele Data CdW HtW WmW Tap Dns Gas Selection / Remarks
54510 Telephone: handset, standard 3 2 ___ yes ___ ___ ___ ___ ___ ___ to bench, optional
9116 Analyser/ tester: electrical safety, automated 3 1 yes opt ___ ___ ___ ___ ___ ___ qty depends on operational requirements; other types of analysers/ testers depend on the warranty of the equipment and the operational requirements
2800 Basin: handwash, Type B 1 1 ___ ___ yes ___ yes yes yes ___
40404 Tapware: sensor 1 1 yes ___ yes yes yes ___ ___ ___ to basin
ID Description Service Category Qty Selection / Remarks
1006 Voice / Data outlet: double Communications 7
5005 Air conditioning: special HVAC positive pressure, filtered
2189 General LED Lighting
6055 Special: task light, built in Lighting under O/H cupboards
7000 Oxygen (O2) Medical Gases 1 for testing/ calibrating equipment
7010 Nitrous Oxide (N2O) Medical Gases 1 for testing/ calibrating equipment
7020 Medical Air (MA) Medical Gases 1 for testing/ calibrating equipment (MA4)
7040 Suction Medical Gases 1 for testing/ calibrating equipment
7070 Scavenge Medical Gases 1 for testing/ calibrating equipment
2577 PO: Double Power 7
11000 Wall panel Services Panels 1 flush mounted
2578 PO: Emergency power, double Power 3
6870 Surgical Air (SA7) Medical Gases 1 for testing/ calibrating equipment
2724 Direct connection: Emergency power Power 1 sensor tap