Mammography - Interventional

Print
Edit it in Health Facility Briefing System (HFBS)
Edit this room data or print a PDF using the HFBS - Health Facility Briefing System. Customise this room data for your project, or add it to an existing health facility brief in HFBS Briefing Module. Edit Briefing
Room Layout Sheet
Download a PDF Room Layout Sheet (RLS) for this standard component. Free View
Other Files
Buy an edible Revit Room Layout Sheet (RLS) for this standard component. $64.99
Online Room Data Sheet
Room Code MAM-INT-I
Net Room Area Briefed 16
Hours of Operation Available 24 Hours
Occupancy 1 patient; Up to 2 staff assisting
Description / Special Requirements X-ray room for mammography imaging procedures which may include interventional procedures such as biopsies; a change area will be provided within the room. Acoustics: Acoustic privacy required Body Protected electrical area Additional Design Considerations: - Radiation shielding as advised by Radiation Consultant - Cable ducts between X-ray table and control area may be required - 'X-ray in Use' lights required outside each entry door
Amendment 24.10.2010 First Draft 21.04.2011 Services modified 24.03.2017 FF & Services updated 03.09.2023 RDS reviewed
ID Fabric Material Finish Specification Selection Remarks
8040 Door protection Composite Prefinished PVC To door frame
4159 Cornice Aluminium Powdercoat Angle flush against wall, in line with ceiling grid
3022 Wall finish Resinate, 6mm Prefinished Splashback to basin & sink, other smooth impervious finish also acceptable
1000 Floor finish Vinyl Standard slip resistant Seamless, coved
2000 Skirting Vinyl Prefinished Floor vinyl coved, 150mmH
8010 Door protection Composite Prefinished PVC Protection plate to 900mm AFFL
5062 Ceiling Acoustic Prefinished Drop-in tiles, 600x600 mm
3000 Wall finish Paint Acrylic, washable
9130 Doors Solid Core Paint 1 1/2 leaf, with lead shielding 1400mm clear opening
ID Description Category Group Qty Selection / Remarks
1050 Bench: laminate Furniture/ Fitting 1 1 with inset sink
2150 Chair: patient Furniture/ Fitting 3 1 to change area
8299 Chair: mammography MEQ - Imaging 3 1 part of mammography system
3400 Coathook Metalwork 1 2 to change area, 1 @1200H
4410 Cupboard: under bench Furniture/ Fitting 1 1 adjustable shelves
4460 Cupboard: wall mounted Furniture/ Fitting 1 1 over bench, adjustable shelves
4700 Curtain track: privacy screen Metalwork 1 1 to change area
5250 Curtain: privacy screen Furniture/ Fitting 3 1 to change area
6152 Dispenser: disposable glove Furniture/ Fitting 2 1 to basin
6154 Dispenser: paper/ hand towel Furniture/ Fitting 2 1 to basin
6158 Dispenser: soap antiseptic Furniture/ Fitting 2 1 to basin
8626 Trolley: dressing, 2 drawer MEQ - Furniture 3 1 under bench
13700 Mirror Metalwork 1 1 to change area
8404 Bracket: sharps disposal MEQ - General 2 1
8318 Suction adapter MEQ - General 3 1
8321 Suction bottle MEQ - General 3 1
8297 Bracket: suction bottle MEQ - General 2 1
8309 Oxygen flowmeter MEQ - General 3 1
8313 Rack: lead apron MEQ - Imaging 1 1
9117 Sharps disposal: wall mounted, 7L MEQ - General 2 1
10061 Waste bin: General, 20 Litres Furniture/ Fitting 3 1
ID Description Grp Qty Ele Data CdW HtW WmW Tap Dns Gas Selection / Remarks
2800 Basin: handwash, Type B 1 1 ___ ___ yes ___ yes yes yes ___
40404 Tapware: sensor 1 1 yes ___ yes yes yes ___ ___ ___ straight spout
40405 Tapware: sink mixer 1 1 ___ ___ yes yes yes ___ ___ ___ provide warm water
47800 Sink, SS: single in bench 1 1 ___ ___ yes opt opt yes yes ___
54510 Telephone: handset, standard 3 1 ___ yes ___ ___ ___ ___ ___ ___ wall mounted
9097 X-ray unit: mammography, 3D 2 1 yes yes ___ ___ ___ ___ ___ ___ with control panel
9092 Control console 1 1 yes yes ___ ___ ___ ___ ___ ___ part of mammography system
ID Description Service Category Qty Selection / Remarks
1006 Voice / Data outlet: double Communications 2 1- patient services panel; 1- mammography control console
5000 Airconditioning HVAC
2242 General LED: colour corrected Lighting
6005 Dimmable Lighting
8010 Staff/ Nurse assist call Nurse Call 1
8080 Emergency call Nurse Call 1
2577 PO: Double Power 1 to wall
2579 PO: Emergency power, single Power 2 to patient services panel
2576 PO: Cleaner Power 1
9040 Body protected Power
12011 Radiation shielding Shielding to walls, floor, ceiling, doors & glazing as specified by Radiation Consultant
2724 Direct connection: Emergency power Power 1 sensor operated tapware
6076 Special: 'X-ray In Use' lights Lighting 1
7000 Oxygen (O2) Medical Gases 1
7040 Suction Medical Gases 1
6838 Trunking: 100mm x 70mm Electrical Installations 1 between generator cabinet, mammography unit, and control console
11505 Floor duct for power & data Services Ducts 1 between generator cabinet, mammography unit, and control console
6827 Cabinet: generator Power 1 for mammography unit; supplied by imaging vendor
2791 PO: Special: 3 phase Power 1 to mammography generator cabinet
2578 PO: Emergency power, double Power 2 to mammography control console
0 0